:SYSTem:ONOFFSync

Syntax

:SYSTem:ONOFFSync {ON|OFF}
:SYSTem:ONOFFSync?

Description

Turn on or off the on/off sync function.
Query whether the on/off sync function is turned on.

Parameter

Name

Type

Range

Default

{ON|OFF}

Bool

ON|OFF

OFF

Explanation

 The specified channels of DP831A, DP831, DP832A and DP832 support the track

function; you can turn on or off the on/off sync function according to your need.

 When the on/off sync function is turned on, the system will track the output

status of the channel; when the on/off sync function is turned off, the system will

not track the output status of the channel.

Return Format

The query returns ON or OFF.

Example

:SYST:ONOFFS ON /*Turn on the on/off sync function*/
:SYST:ONOFFS? /*Query whether the on/off sync function is turned on and

the query returns ON*/

Related

Command

:SYSTem:TRACKMode

:SYSTem:OTP

Syntax

:SYSTem:OTP {ON|OFF}
:SYSTem:OTP?

Description

Enable or disable the over-temperature protection (OTP) function.
Query the status of the over-temperature protection function.

Parameter

Name

Type

Range

Default

{ON|OFF}

Bool

ON|OFF

ON

Explanation

When the OTP function is enabled, the instrument turns off the output

automatically when the temperature inside the instrument reaches the limit.

Return Format

The query returns ON or OFF.

Example

:SYST:OTP ON /*Enable the OTP function*/
:SYST:OTP? /*Query the status of the OTP function and the query returns ON*/