Trigger:in[:enable, Trigger:in[:enable] -145 – RIGOL DP832A User Manual

Page 169

Advertising
background image

Chapter 2 Command System

RIGOL

DP800 Programming Guide

2-145

:TRIGger:IN[:ENABle]

Syntax

:TRIGger:IN[:ENABle] [D0|D1|D2|D3,]{ON|OFF}
:TRIGger:IN[:ENABle]? [D0|D1|D2|D3]

Description

Enable or disable the trigger input function of the specified data line.
Query the status of the trigger input function of the specified data line.

Parameter

Name

Type

Range

Default

[D0|D1|D2|D3]

Discrete

D0|D1|D2|D3

D0

{ON|OFF}

Bool

ON|OFF

OFF

Explanation

 If [D0|D1|D2|D3] is omitted, the command sets the trigger input function of

the data line currently selected.

 After enabling the trigger input function, the specified source under control

(

:TRIGger:IN:SOURce

) will turn on the output, turn off the output or toggle the

output state according to the setting of the

:TRIGger:IN:RESPonse

command

when the input signal on the specified data line meets the current trigger type

(

:TRIGger:IN:TYPE

).

Return Format

The query returns Dn,ON or Dn,OFF; wherein, n=0, 1, 2 or 3.

Example

:TRIG:IN D1,ON /*Enable the trigger input function of D1*/
:TRIG:IN? D1 /*Query the status of the trigger input function of D1 and the

query returns D1,ON*/

Related

Commands

:TRIGger:IN:TYPE

:TRIGger:IN:RESPonse

:TRIGger:IN:SOURce

Advertising