Trigger:in:chtype, Trigger:in:chtype -143 – RIGOL DP832A User Manual

Page 167

Advertising
background image

Chapter 2 Command System

RIGOL

DP800 Programming Guide

2-143

:TRIGger:IN:CHTYpe

Syntax

:TRIGger:IN:CHTYpe {BUS|IMM}

:TRIGger:IN:CHTYpe?

Description

Select the trigger source type.
Query the trigger source type currently selected.

Parameter

Name

Type

Range

Default

{BUS|IMM}

Discrete

BUS|IMM

BUS

Explanation

 The trigger source types include BUS (bus trigger) and IMM (immediate trigger).

For the bus trigger (BUS; also called software trigger), the power supply

receives trigger from the bus; namely, the power supply receives trigger via

software. For immediate trigger (IMM), the power supply receives immediate

trigger from the remote interface.

 When the trigger source type is "IMM" (immediate trigger), the power supply

will execute a complete trigger operation (the voltage/current of the specified

channel changes to the trigger voltage/current currently set immediately)

immediately without any delay when you execute the

:INITiate

or

:TRIGger:IN:IMMEdiate

command.

 When the trigger source type is "BUS" (bus trigger; namely software trigger),

after you execute the

:INITiate

or

:TRIGger:IN:IMMEdiate

command to initialize

the trigger system, the power supply generates a trigger when you execute the

*TRG

command and then executes a trigger operation (the voltage/current of

the specified channel changes to the trigger voltage/current currently set) after

waiting for the specified delay time (if you have set the delay time; refer to

the

:TRIGger[:SEQuence]:DELay

command).

 When the trigger source type is "BUS" (bus trigger; namely software trigger),

you can ensure the synchronization of the operations by executing the

*WAI

command. After you execute the

*WAI

command, the power supply will only

execute the new command after all the previous operations are finished.

 When the trigger source type is "BUS" (bus trigger; namely software trigger),

you can make the system report the completion of the operation by executing

the

*OPC

command. When you execute the *OPC? command, the system will

return "1" to the output buffer when the operation is completed; when you

execute the *OPC command, the system will set bit0 (OPC bit; operation

completed) in the standard event register when the operation is completed.

 You can also send the

:TRIGger[:SEQuence]:SOURce

command to select the

trigger source type.

Return Format

The query returns BUS or IMM.

Example

:TRIG:IN:CHTY BUS /*Set the trigger source type to BUS (bus trigger; namely

software trigger)*/

:TRIG:IN:CHTY? /*Query the trigger source type currently selected and the

query returns BUS*/

Related

Commands

:INITiate

:TRIGger:IN:IMMEdiate

*TRG

:TRIGger[:SEQuence]:DELay
*WAI

Advertising