Trigger:out[:enable, Trigger:out[:enable] -153 – RIGOL DP832A User Manual

Page 177

Advertising
background image

Chapter 2 Command System

RIGOL

DP800 Programming Guide

2-153

:TRIGger:OUT[:ENABle]

Syntax

:TRIGger:OUT[:ENABle] [D0|D1|D2|D3,]{ON|OFF}
:TRIGger:OUT[:ENABle]? [D0|D1|D2|D3]

Description

Enable or disable the trigger output function of the specified data line.
Query the status of the trigger output function of the specified data line.

Parameter

Name

Type

Range

Default

[D0|D1|D2|D3]

Discrete

D0|D1|D2|D3

D0

{ON|OFF}

Bool

ON|OFF

OFF

Explanation

 If [D0|D1|D2|D3] is omitted, the command enables or disables the trigger

output function of the data line currently selected.

 When the trigger output function is enabled, the specified data line outputs the

specified level or square waveform according to the setting of the output signal

(

:TRIGger:OUT:SIGNal

) when the output signal of the specified control source

(

:TRIGger:OUT:SOURce

) meets the trigger condition set

(

:TRIGger:OUT:CONDition

).

Return Format

The query returns Dn,ON or Dn,OFF; wherein, n=0, 1, 2 or 3.

Example

:TRIG:IN D1,ON /*Enable the trigger output function of D1*/

:TRIG:IN? D1 /*Query the status of the trigger output function of D1 and the

query returns D1,ON*/

Related

Commands

:TRIGger:OUT:SOURce
:TRIGger:OUT:CONDition
:TRIGger:OUT:SIGNal

Advertising
This manual is related to the following products: