Output:ocp:clear, Output:ocp:clear -55 – RIGOL DP832A User Manual

Page 79

Advertising
background image

Chapter 2 Command System

RIGOL

DP800 Programming Guide

2-55

:OUTPut:OCP:CLEAR

Syntax

:OUTPut:OCP:CLEAR [CH1|CH2|CH3]

Description

Clear the label of the overcurrent protection occurred on the specified channel.

Parameter

Name

Type

Range

Default

[CH1|CH2|CH3]

Discrete

CH1|CH2|CH3

None

Explanation

 Before executing the command, make sure that the reason that causes the

OCP on the specified channel is cleared (you can decrease the output current

to be below the OCP value or increase the OCP value to be greater than the

output current).

 When [CH1|CH2|CH3] is omitted, the system clears the label of the OCP

occurred on the current channel.

 You can also send the

[:SOURce[<n>]]:CURRent:PROTection:CLEar

command

to clear the circuit and label of the OCP occured on the specified channel and

turn on the corresponding channel output.

 You can send the

:OUTPut:OCP:ALAR?

or

:OUTPut:OCP:QUES?

command to

query whether OCP occurred on the specified channel.

Example

:OUTP:OCP:QUES? CH1 /*Query whether overcurrent protection currently

occurred on CH1 and the query returns YES*/

(Operation: decrease the output current to be below the OCP value or increase the

OCP value to be greater than the output current)
:OUTP:OCP:CLEAR CH1 /*Clear the label of the overcurrent protection occurred

on CH1*/

:OUTP:OCP:QUES? CH1 /*Query whether overcurrent protection occurred on

CH1 and the query returns NO*/

Related

Commands

:OUTPut:OCP:ALAR?
:OUTPut:OCP:QUES?

[:SOURce[<n>]]:CURRent:PROTection:CLEar

Advertising