Analyzer:object, Analyzer:result, Analyzer:object -6 – RIGOL DP832A User Manual

Page 30: Analyzer:result? -6

Advertising
background image

RIGOL

Chapter 2 Command System

2-6

DP800 Programming Guide

:ANALyzer:OBJect

Syntax

:ANALyzer:OBJect {V|C|P}
:ANALyzer:OBJect?

Description

Set the analysis object of the analyzer to voltage, current or power.
Query the analysis object of the analyzer.

Parameter

Name

Type

Range

Default

{V|C|P}

Discrete

V|C|P

V

Explanation

You can only set the analysis object when valid record file is opened (refer to

the

:ANALyzer:FILE?

command).

Return Format

The query returns V, C or P.

Example

:ANAL:OBJ V /*Set the analysis object of the analyzer to voltage*/
:ANAL:OBJ? /*Query the analysis object of the analyzer and the query

returns V*/

Related

Command

:ANALyzer:FILE?

:ANALyzer:RESult?

Syntax

:ANALyzer:RESult?

Description

Query the analysis results, including the number of groups, median, mode, average,

variance, range, minimum, maximum and mean deviation.

Return Format

The query returns the analysis results with the data separated by commas, for

example,

Group:85,Median:41.9994V,Mode:0.0000V,Average:34.0924V,Variance:269.5170V,

Range:42.0002V,Min:0.0000V,Max:42.0002V,Mean:12.8347V.

Example

:ANAL:RES? /*Query the analysis results and the query returns

Group:85,Median:41.9994V,Mode:0.0000V,Average:34.0924V,Va

riance:269.5170V,Range:42.0002V,Min:0.0000V,Max:42.0002V,M

ean:12.8347V*/

Related

Command

:ANALyzer:ANALyze

Advertising
This manual is related to the following products: