Output:ocp:value, Output:ocp:value -57 – RIGOL DP832A User Manual

Page 81

Advertising
background image

Chapter 2 Command System

RIGOL

DP800 Programming Guide

2-57

:OUTPut:OCP:VALue

Syntax

:OUTPut:OCP:VALue [CH1|CH2|CH3,]{<value>|MINimum|MAXimum}
:OUTPut:OCP:VALue? {CH1|CH2|CH3}[,MINimum|MAXimum]
:OUTPut:OCP:VALue? [MINimum|MAXimum]

Description

Set the overcurrent protection value of the specified channel.
Query the overcurrent protection value of the specified channel.
Query the overcurrent protection value of the current channel.

Parameter

Name

Type

Range

Default

[CH1|CH2|CH3]

Discrete

CH1|CH2|CH3

None

<value>

Real

Refer to Table 2-2

{CH1|CH2|CH3}

Discrete

CH1|CH2|CH3

None

Explanation

 When the overcurrent protection function is enabled, the output will turn off

automatically when the output current exceeds the overcurrent protection

value currently set. You can send the

:OUTPut:OCP:ALAR?

or

:OUTPut:OCP:QUES?

command to query whether overcurrent protection

occurred on the specified channel currently.

 When [CH1|CH2|CH3] is omitted, the system performs the corresponding

operation on the current channel.

 You can also send the

[:SOURce[<n>]]:CURRent:PROTection[:LEVel]

command to set the OCP value of the specified channel.

Return Format

The query returns the overcurrent protection value, for example, 5.0000.

Example

:OUTP:OCP:VAL CH1,5 /*Set the OCP value of CH1 to 5A*/
:OUTP:OCP:VAL? CH1 /*Query the OCP value of CH1 and the query returns

5.0000*/

Related

Commands

:OUTPut:OCP:ALAR?
:OUTPut:OCP:QUES?
[:SOURce[<n>]]:CURRent:PROTection[:LEVel]