Timer:endstate, Timer:endstate -130 – RIGOL DP832A User Manual

Page 154

Advertising
background image

RIGOL

Chapter 2 Command System

2-130

DP800 Programming Guide

:TIMEr:ENDState

Syntax

:TIMEr:ENDState {OFF|LAST}
:TIMEr:ENDState?

Description

Set the end state of the timer.
Query the current end state of the timer.

Parameter

Name

Type

Range

Default

{OFF|LAST}

Discrete

OFF|LAST

OFF

Explanation

 The end state refers to the state of the instrument after it finishes outputting

the total number of groups of voltage/current values when the number of

cycles is a finite value.

 OFF: the instrument turns off the output automatically after finishing the

output.

 LAST: the instrument stops at the output state of the last group after finishing

the output.

 The total number of groups in timing output = the number of groups × the

number of cycles; wherein, the number of groups and the number of cycles are

set by the

:TIMEr:GROUPs

and

:TIMEr:CYCLEs

commands respectively.

Return Format

The query returns OFF or LAST.

Example

:TIME:ENDS LAST /*Set the end state of the timer to "Last"*/
:TIME:ENDS? /*Query the current end state of the timer and the query

returns LAST*/

Related

Commands

:TIMEr:CYCLEs
:TIMEr:GROUPs