Display[:window][:state, Display[:window]:text:clear, Display[:window][:state] -20 – RIGOL DP832A User Manual

Page 44: Display[:window]:text:clear -20

Advertising
background image

RIGOL

Chapter 2 Command System

2-20

DP800 Programming Guide

:DISPlay[:WINDow][:STATe]

Syntax

:DISPlay[:WINDow][:STATe] {OFF|ON}

:DISPlay[:WINDow][:STATe]?

Description

Turn off or on the screen display.
Query the current screen display state.

Parameter

Name

Type

Range

Default

{OFF|ON}

Bool

OFF|ON

ON

Explanation

 OFF: turn off the screen display; ON: turn on the screen display.

 Turning off the screen display is only valid in remote mode. When the

instrument returns to the local mode, the screen display is turned on
automatically. Pressing Back can return the instrument from remote mode to
local mode.

Return Format

The query returns OFF or ON.

Example

:DISP OFF /*Turn off the screen display*/
:DISP? /*Query the current screen display state and the query returns

OFF*/

:DISPlay[:WINDow]:TEXT:CLEar

Syntax

:DISPlay[:WINDow]:TEXT:CLEar

Description

Clear the characters displayed on the screen.

Example

:DISP:TEXT:CLE /*Clear the characters displayed on the screen*/

Related

Command

:DISPlay[:WINDow]:TEXT[:DATA]

Advertising
This manual is related to the following products: