Chapter 2 Command System

RIGOL

DP800 Programming Guide

2-131

:TIMEr:GROUPs

Syntax

:TIMEr:GROUPs <value>
:TIMEr:GROUPs?

Description

Set the number of output groups of the timer.
Query the current number of output groups of the timer.

Parameter

Name

Type

Range

Default

<value>

Integer

1 to 2048

1

Explanation

 The number of output groups is defined as the number of groups of preset

voltage/current values that the power supply outputs in each cycle.

 The total number of groups in timing output = the number of groups × the

number of cycles; wherein, the number of cycles is set by the

:TIMEr:CYCLEs

command.

 The power supply will terminate the timer function when the total number of

groups of outputs is finished. At this point, the state of the power supply

depends on the setting of the

:TIMEr:ENDState

command.

Return Format

The query returns an integer from 1 to 2048, for example, 25.

Example

:TIME:GROUP 25 /*Set the number of output groups of the timer to 25*/
:TIME:GROUP? /*Query the current number of output groups of the timer and

the query returns 25*/

Related

Commands

:TIMEr:CYCLEs
:TIMEr:ENDState