Timer:parameter, Timer:parameter -132 – RIGOL DP832A User Manual

Page 156

Advertising
background image

RIGOL

Chapter 2 Command System

2-132

DP800 Programming Guide

:TIMEr:PARAmeter

Syntax

:TIMEr:PARAmeter <secnum>,<volt>,<curr>,<time>
:TIMEr:PARAmeter? <firnum>[,<timercount>]

Description

Set the timer parameters of the specified group.
Query the timer parameters of the specified groups.

Parameter

Name

Type

Range

Default

<secnum>

Integer

0 to 2047

None

<volt>

Real

The voltage range of the

current channel

CH1/CH2: 1V

CH3: -1V

<curr>

Real

The current range of the

current channel

1A

<time>

Real

1s to 99999s

1s

<firnum>

Integer

0 to 2047

None

<timercount>

Integer

1 to 2048

1

Explanation

 <secnum> is the group number of the specified group of timer parameters;

<volt>, <curr> and <time> are the voltage, current and time of the specified

group of timer parameters respectively and their units are V, A and s

respectively.

 The query command queries the specified groups of timer parameters (the

numbers of the groups are continuous). <firnum> is the group number of the

first group of timer parameters to be queried. <timercount> is the total

number of groups of timer parameters to be queried.

 When <timercount> is omitted, the command queries a single group of timer

parameters by default.

 You can also send the

:OUTPut:TIMEr

command to set or query the timer

parameters.

Return Format

The query returns a string starting with #.

For example, #90000000361,8.000,1.0000,10;2,6.000,1.0000,10;;

wherein, #9000000036 is the data block header;

1,8.000,1.0000,10;2,6.000,1.0000,10; are the specified timer parameters.

 The data block header is used to describe the length information of the data

stream and starts with #. For example, the number "9" in #9000000036

denotes that the 9-bit data (000000036) following it is used to denote the data

stream length (36 bytes).

 The format of each group of timer parameters is "number,voltage,current,time"

and multiple groups of timer parameters are separated by ";". For example,

1,8.000,1.0000,10;2,6.000,1.0000,10; denotes two groups of timer

parameters; the number of the first group of timer parameters is 1, the voltage

is 8.000V, the current is 1.0000A and the time is 10s; the number of the second

group of timer parameters is 2, the voltage is 6.000V, the current is 1.0000A

and the time is 10s.

Example

:TIME:PARA 1,8,1,10 /*Set the timer parameters of the first group to 8V, 1A, 10s*/
:TIME:PARA 2,6,1,10 /*Set the timer parameters of the second group to 6V, 1A,

10s*/

:TIME:PARA? 1,2 /*Query two groups of timer parameters starting from the

first group and the query returns

#90000000361,8.000,1.0000,10;2,6.000,1.0000,10;*/