Output commands, Output commands -52 – RIGOL DP832A User Manual

Page 76

Advertising
background image

RIGOL

Chapter 2 Command System

2-52

DP800 Programming Guide

:OUTPut Commands

The :OUTPut commands are used to enable or disable the channel output, overvoltage/overcurrent

protection functions, track function and Sense function, query the output mode of the channel as well as set

and query the related information of overvoltage/overcurrent protection. For single-channel model, you can

also select and query the current range of the channel. The range and default value of

overvoltage/overcurrent protection corresponding to each channel (multi-channel models) or range

(single-channel model) of different models are as shown in the table below.

Table 2-2 Range and default value of overvoltage/overcurrent protection of each channel (range) of

different models of DP800 series

Channel (Range)

OVP/OCP Settable Range

OVP/OCP Default Value

DP831A

CH1 (8V/5A)

CH2 (30V/2A)

CH3 (-30V/2A)

1mV to 8.8V/0.1mA to 5.5A

1mV to 33V/0.1mA to 2.2A

-1mV to -33V/0.1mA to 2.2A

8.800V/5.5000A

33.000V/2.2000A

-33.000V/2.2000A

DP832A

CH1 (30V/3A)

CH2 (30V/3A)

CH3 (5V/3A)

1mV to 33V/1mA to 3.3A

1mV to 33V/1mA to 3.3A

1mV to 5.5V/1mA to 3.3A

33.000V/3.300A

33.000V/3.300A

5.500V/3.300A

DP821A

CH1 (60V/1A)

CH2 (8V/10A)

1mV to 66V/0.1mA to 1.1A

1mV to 8.8V/1mA to 11A

66.000V/1.1000A

8.800V/11.000A

DP811A

Range1 (20V/10A)

Range2 (40V/5A)

1mV to 22V/0.1mA to 11A

1mV to 44V/0.1mA to 5.5A

22.000V/11.0000A

DP831

[1]

CH1 (8V/5A)

CH2 (30V/2A)

CH3 (-30V/2A)

10mV to 8.8V/1mA to 5.5A

10mV to 33V/1mA to 2.2A

-10mV to -33V/1mA to 2.2A

8.800V/5.500A

33.00V/2.200A

-33.00V/2.200A

DP832

[1]

CH1 (30V/3A)

CH2 (30V/3A)

CH3 (5V/3A)

10mV to 33V/1mA to 3.3A

10mV to 33V/1mA to 3.3A

10mV to 5.5V/1mA to 3.3A

33.00V/3.300A

33.00V/3.300A

5.50V/3.300A

DP821

[1]

CH1 (60V/1A)

CH2 (8V/10A)

10mV to 66V/10mA to 1.1A

10mV to 8.8V/10mA to 11A

66.00V/1.100A

8.80V/11.00A

DP811

[1]

Range1 (20V/10A)

Range2 (40V/5A)

10mV to 22V/10mA to 11A

10mV to 44V/10mA to 5.5A

22.00V/11.00A

Note

[1]

: When DP831 (DP832, DP821 or DP811) is installed with the high resolution option, its overvoltage/overcurrent

protection settable range and default value of each channel are the same with those of DP831A (DP832A, DP821A or

DP811A).


Command List:

:OUTPut:CVCC?

:OUTPut:MODE?

:OUTPut:OCP:ALAR?

:OUTPut:OCP:QUES?

:OUTPut:OCP:CLEAR

:OUTPut:OCP[:STATe]

:OUTPut:OCP:VALue

Advertising
This manual is related to the following products: