Chapter 2 Command System

RIGOL

DP800 Programming Guide

2-155

:TRIGger:OUT:POLArity

Syntax

:TRIGger:OUT:POLArity [D0|D1|D2|D3,]{POSItive|NEGAtive}
:TRIGger:OUT:POLArity? [D0|D1|D2|D3]

Description

Set the polarity of the trigger output signal of the specified data line.
Query the polarity of the trigger output signal of the specified data line.

Parameter

Name

Type

Range

Default

[D0|D1|D2|D3]

Discrete

D0|D1|D2|D3

D0

{POSItive|NEGAtive}

Discrete

POSItive|NEGAtive

POSItive

Explanation

 If [D0|D1|D2|D3] is omitted, the command sets the polarity of the trigger

output signal of the data line currently selected.

 POSItive: output the currently specified output signal

(

:TRIGger:OUT:SIGNal

,

:TRIGger:OUT:DUTY

and

:TRIGger:OUT:PERIod

)

when the trigger condition is met.

NEGAtive: turn the currently specified output signal

(

:TRIGger:OUT:SIGNal

,

:TRIGger:OUT:DUTY

and

:TRIGger:OUT:PERIod

)

upside down and then output the signal when the trigger condition is met.

Return Format

The query returns POSITIVE or NEGATIVE.

Example

:TRIG:OUT:POLA D1,NEGAtive /* Set the polarity of the trigger output signal of

D1 to negative */

:TRIG:OUT:POLA? D1 /* Query the polarity of the trigger output signal of

D1 and the query returns NEGATIVE*/

Related

Commands

:TRIGger:OUT:SIGNal

:TRIGger:OUT:DUTY

:TRIGger:OUT:PERIod