Trigger:in:voltage, Trigger:in:voltage -150 – RIGOL DP832A User Manual

Page 174

Advertising
background image

RIGOL

Chapter 2 Command System

2-150

DP800 Programming Guide

:TRIGger:IN:VOLTage

Syntax

:TRIGger:IN:VOLTage {CH1|CH2|CH3},<voltage>

Description

Set the trigger voltage of the specified channel.

Parameter

Name

Type

Range

Default

{CH1|CH2|CH3}

Discrete

CH1|CH2|CH3

None

<voltage>

Real

0 to the maximum voltage

of the specified channel

0V

Explanation

 When the trigger source type (refer to the

:TRIGger[:SEQuence]:SOURce

or

:TRIGger:IN:CHTYpe

command) is set to "IMM" (immediate trigger), the

power supply will execute a complete trigger operation (the voltage/current of

the specified channel changes to the trigger voltage/current currently set

immediately) immediately when you execute the

:INITiate

or

:TRIGger:IN:IMMEdiate

command.

 When the trigger source type (refer to the

:TRIGger[:SEQuence]:SOURce

or

:TRIGger:IN:CHTYpe

command) is set to "BUS" (bus trigger; namely software

trigger), after you execute the

:INITiate

or

:TRIGger:IN:IMMEdiate

command to

initialize the trigger system, the power supply generates a trigger when you

execute the

*TRG

command and then executes a trigger operation (the

voltage/current of the specified channel changes to the trigger voltage/current

currently set) after waiting for the specified delay time (if you have set the delay

time; refer to the

:TRIGger[:SEQuence]:DELay

command).

 You can also send the

[:SOURce[<n>]]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]

command to set the

trigger voltage of the specified channel.

Example

:TRIG:IN:VOLT CH1,1 /*Set the trigger voltage of CH1 to 1V*/

Related

Commands

:TRIGger[:SEQuence]:SOURce

:TRIGger:IN:CHTYpe

:INITiate

:TRIGger:IN:IMMEdiate

*TRG
:TRIGger[:SEQuence]:DELay

[:SOURce[<n>]]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]

Advertising
This manual is related to the following products: