Specifications, 1 specifications, 1 general specifications – Yokogawa GD40 Gas Density Detector User Manual

Page 10: Specifications -1, Specifications -1 1.1.1, General specifications -1

Advertising
background image

<1. Specifications>

1-1

IM 11T3E1-01E

1.

SPECIFICATIONS

1.1

Specifications

The specifications of the GD402 gas density meter as described below.

Certificate is approved for model GD402G, GD40G.

1.1.1

General Specifications

l

System Components

(1) GD40G, T, V, R Detector:Rainproof for outdoor use (equivalent to IP65/NEMA4X)

(see note under “1.1.2 Ambient conditions” on page1-5.)

Ambient Temperature :

-10 to 60°C

Ambient Humidity :

5 to 95%RH

GD40G : General purpose detector.

(Non Explosion-proof)

Electrical connection : 1/2 NPT female

Process connection : 1/4 NPT female

GD40T : FM Explosion-proof and Intrinsically safe Approval.

Explosion-proof for Class I, Division 1, Groups B, C and D;

Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically

Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.

Enclosure : NEMA Type 4X

Temperature Code : T5

Electrical connection : 1/2 NPT female

Process connection : 1/4 NPT female

GD40V : CSA Explosion-proof and Intrinsically safe Approval.

Explosion-proof for Class I, Division 1, Groups B, C and D;

Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically

Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.

Enclosure :

Type 4X

Temperature Code : T5

Electrical connection : 1/2 NPT female

Process connection : 1/4 NPT female

GD40R : TIIS Explosion-proof and Intrinsically safe Approval.

Explosion-proof code : Exd [ia] IIB+H

2

T5

Temperature Code : T5

Electrical connection : G1/2 female

Process connection : Rc1/4

Advertising