3 cables wired to outputs, Cables wired to outputs -17 – Yokogawa GD40 Gas Density Detector User Manual

Page 40

Advertising
background image

<2. Installation, Wiring and Piping>

2-17

IM 11T3E1-01E

100 to 240 V AC

50/60Hz

Surge Protector

GD402 Converter

100-240 V AC type

SUPPLY

L N G

Green cable

1m

max

External circuit breaker

Rating: 5 A

Case-grounding

terminal

Case-grounding

terminal

SUPPLY

+ - G

24 V DC

F0216.ai

GD402 Converter

24 V DC type

Use IEC947-1 or IEC947-3 compliant product.

Terminal box

*

1

*

1 Surge protector and terminal box should be installed in order to meet

the requirements of CE marking.

*

1

Figure 2.12 Cables Wired to Power Supply

2.4.3

Cables Wired to Outputs

These cables are used to transmit 4-20 mA DC signals and carry out BRAIN communication.

Use shielded cables of 8 to 16 mm in finished outer diameter and 0.75 mm

2

minimum in thick-

ness (or a two-core shielded cable for single output).
Wire the cables as instructed below:
(1) Use cables that are 0.75 to 2.5 mm

2

thick. Treat the cable ends by stripping the core wires

back 7 mm (see Figure 2.14).

(2) Wire the cable for output 1 to terminals 3 and 4 on the block and the cable for output 2 to

terminals 5 and 6 on the block. Beware of the correct polarities. Note that only terminals 3 and

4 (output 1) are effective for BRAIN communication.

Loosen the terminal screws, insert the stripped ends of the core wires into the terminals, and

fasten the screws to fix the cables.

(3) Ground the shielding wire at a given terminal on the converter; do not ground the other end of

the cable. (The appropriate tightening torque for the internal wiring terminals is 0.4 N·m.)

7mm

F0217.ai

Figure 2.13 Examples of Treatment on Cable End Wired to the Converter