<5. DENSITY / PARAMETER SETTING>

5-17

IM 11T3E1-01E

CODE 01: Setting of Hold during Maintenance

This setting determines whether or not the output signal is held during maintenance. As a value

to be held, you can select either a value immediately before or a preset value.

(a) Hold enable/disable [*M_HLD]
・ Disable:

0

・ Enable:

1

・ Enable (preset value): 2
(b) Setting of preset hold value [*PR.SET]
This item applies if item (a) above is set to "2".
Configurable range:

-10.0 to 110.0 (percent of analog output)

[ENT]

01

*CODE

*SERVC

X

*M_HLD

XXX.X

*PR.SET

[ENT]

[ENT]

[YES]

F0508.ai

*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.

*2: Enter the code number.

*3: Set a value using the [ ] and [ENT] keys.

*4: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].

*

1

*

2

*

3

*

4

*

5

Figure 5.7 Setting of Hold during Maintenance