<2. Installation, Wiring and Piping>

2-19

IM 11T3E1-01E

2.4.5

Cable Wired to GD40 Detector

The cable wired to the GD40 detector comes from the GD402 converter. Use a shielded cable

of 10 to 13.5 mm in finished outer diameter that suits the dedicated cable gland. Wire the cable

through the cable gland, being careful not to connect the cable with the wrong polarities. Attach

M4 crimp terminals to the ends of the core wires (see Figure 2.14). Connect the shielding wire to

terminal 13 on the converter (see Figure 2.12). When crimping the terminals, use tools that fit the

terminals.
If a malfunction occurs and it is assumed to be due to noise, use a double-shielded cable (for ex-

ample, the cable with Yokogawa’s Model GDW-L  ). Connect the outer shielding wire to ter-

minal 13 on the converter and the other end of the wire to the grounding terminal on the detector.

Connect the inner shielding wires to the grounding terminal on the converter (see Figure 2.17).

Terminals for

wiring to converter

F0219.ai

Class A

grounding terminal

Figure 2.15 Configuration of Detector Terminals

11 12 13

Outer shielding wire

Class A

grounding

terminal

Do not connect the inner

shielding wires here.

Inner shielding

wires

Outer shielding

wire

GD40T, V, R

Detector

GD402G, T, V, R

Converter

F0220.ai

Class A

grounding

Figure 2.16 Connection Between Detector and Converter When Using a Double-Shielded Cable

Advertising