<2. Installation, Wiring and Piping>

2-14

IM 11T3E1-01E

Contact input

MAINT

ALM

FAIL

SPAN/FUNC

ZERO/SEL GAS

CONT

INP

FAIL

Contact output

ALARM

Contact output

MAINTENANCE

Contact output

FUNCTION

Contact output

SELECT GAS

Contact output

Power supply

100 to 240 V AC

or 24 V DC

Case grounding

terminal

Case grounding

terminal

GD402 Converter

Isolated 4-20 mA Output

with BRAIN Communication

Isolated 4-20 mA Output

Pressure Transmitter

+

-

+

-

+

-

+

-

-

+

+

L or +

-

N or -

+

SHIELD

-

GD40 Detector

14
15

16
17

18

19
20
21

22
23
24

25

1
2

3
4

5

6

7

8
9

10

11

12

13

26

ANLG

OUT1

ANLG

OUT2

SNSR

PWR

SNSR

INP

DET

INP

-

+

Case grounding

terminal

Surge protector

F0215.ai

*

1

Intrinsic safety

Grounding

Class A grounding

*

2

*

4

Terminal

Indication

Shield

Requirement

Requirement

MAINTENANCE

Contact output

MAINT

Unshielded

ALARM

Contact output

ALM

Unshielded

FAIL

Contact output

FAIL

Unshielded

FUNCTION

Contact output

SPAN

Unshielded

SELECT GAS

Contact output

ZERO

Unshielded

Contact input

CONT IN

Shielded

Analog output1

ANLG OUT1

Shielded

Analog output2

ANLG OUT2

Shielded

Pressure

transmitter input

Detector input

SNSR PWR

SNSR INP

Shielded

DET INP

SHIELD

Shielded

Supply

L, N, G

Unshielded

For outdoor wiring, cable length should be less than 30 m .

Shield should be grounded at one end only.

*

3

Total resistance should not exceed 50 Ω.

For outdoor wiring, cable length should be less than 30 m .

Shield should be grounded at one end only.

Maximum load resistance including wire resistance is 600Ω . When

BRAIN communication is used, it is 250 to 550 Ω.

*

3

Total resistance should not exceed 50 Ω.

For outdoor wiring, cable length should be less than 30 m

.

Shield should be grounded at one end only.

*

3

Total resistance should not exceed 50 Ω.

For outdoor wiring, cable length should be less than 30 m

.

Connect shield to SHIELD terminal on converter.

*

3

Cable List

*1 Surge protector is not provided with GD402T/V/R type converter.

*2 Intrinsic safety grounding

GD402V, GD40V; All wiring should comply with Canadian Electrical Code and Local Electrical Codes.

GD402T, GD40T; All wiring should comply with National Electrical Code and ANSI/NFPA 70 and

Local Electrical Codes.

*3 Restriction on cable length does not apply to GD402T/V/R type converter.

*4 When select GD402T/V/R.

(grounding resistance

10 Ω or less)

Figure 2.11 Terminals on the Converter