<5. DENSITY / PARAMETER SETTING>

5-6

IM 11T3E1-01E

Table 5.4 Setting Parameters at Service Level

Code

No.

Item

Display

Data to be Set

(or Conditions)

Remarks

20 Selection of pressure unit

*PRES.U

· kPa : 0

· MPa : 1

· psi : 2

See Figure 5.17 for key operation.

21 Selection of density unit

*DENS.U

· kg/m

3

: 0

· lb/ft

3

: 1

See Figure 5.18 for key operation.

22 Selection of calorific value unit

*CAL.U

· MJ/m

3

: 0

· kBTU/ft

3

: 1

See Figure 5.19 for key operation.

23 Selection of temperature unit

*TEMP.U

· °C : 0

· °F : 1

See Figure 5.20 for key operation.

30 Setting of calendar

Setting of year/month/day

Setting of hour/minute

*Y_M_D

*H_M

00.01.01 to 99.12.31

00.00 to 23.59

See Figure 5.21 for key operation.

For leap years, the setting can

cope with ‘February 29’.

31 Showing/hiding of negative measured

values (-)

*MINUS

· Show : 0

· Hide : 1

See Figure 5.22 for key operation.

40 Indication of calibration coefficients

(read-only)

Zero

Span

*C_K_Z

*C_K_S

SX.XXXX

X.XXXX

See Figure 5.23 for key operation.

S of SX.XXXX means Marks such

as a plus or the minus.

41 Indication of oscillation frequency

(read-only)

Frequency F2

Frequency F4

Frequency ratio

*F2.KHZ

*F4.KHZ

*F2/F4

X.XXXXX

X.XXXXX

X.XXXXX

See Figure 5.24 for key operation.

42 Indication of software version

(read-only)

*REV

X.XX

See Figure 5.25 for key operation.

43 Setting of high-resolution mode

*S_CYC

· Normal :

0

· High-resolution : 1

See Figure 5.26 for key operation.

44 Setting of password

*PASS

0.0.0 to 9.9.9

See Figure 5.27 for key operation.

45 Selecting Battery alarm detection

*BAT

· Non detection : 0

· Detection : 1

See Figure 5.28 for key operation.

50 Selecting the instruments

*MODEL

· Density meter : 0

· Calorie meter : 1

· Hydrogen purity, replacement

meter : 2

See Figure 5.29 for key operation.

82 Setting Detector Constant

See Figure 5.30 for key operation.

5.2

Parameter Setting

[Aborting the Setting Operation]
Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement

(<MEASURE>) mode to a mode at the operation level. It is also used to return to the measure-

ment mode from other modes.

5.2.1

Parameter Setting at Operation Level

To move to the operation level, press the [MODE] key and then enter the password (XXX). Pass-

word is not selected in first condition. When password is necessary, refer to Section 4.2 “Setting

Lists” and selected password what you need.