Code 20: selection of pressure unit [ *pres.u – Yokogawa GD40 Gas Density Detector User Manual

Page 146

Advertising
background image

<7. HYDROGEN PURITY / PARAMETER SETTING>

7-23

IM 11T3E1-01E

CODE 14: Setting for Remote Semi-automatic Calibration [*REMOT]

This setting determines whether or not the measuring range of the replacement (substituents-

concentration) meter is selected remotely.

If remote range selection is to be carried out, you must wire the contact input.

・ Disable:

0

・ Enable:

1

If the code is set to “1”, the analyzer selects:
• the “AIR in CO

2

” range, if the contact input is open.

• the “H

2

in CO

2

” range, if the contact input is closed.

[ENT]

*CODE

14

*SERVC

X

*REMOT

[ENT]

[YES]

F0716.ai

*

1

*

2

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

*

3

Figure 7.16 Setting for Remote Semi-automatic Calibration

CODE 20: Selection of Pressure Unit [*PRES.U]

This setting defines the pressure unit for the gas pressure transmitter.

・ kPa: 0
・ MPa: 1
・ psi :

2

[ENT]

*CODE

20

*SERVC

X

*PRES.U

[ENT]

[YES]

F0717.ai

*

1

*

2

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

*

3

Figure 7.17 Selection of Pressure Unit