<3. Operation>

3-7

IM 11T3E1-01E

3.1.7

Checking the Analyzer for Performance

When the required parameters have been completely set, bring into operation all equipment

composing the measurement loop. Keep this equipment operating for a while. After making sure

there is no problem with the equipment, go into normal operation.
For reference, the following summarizes the setting functions (related to signals) of the GD402

analyzer that are used to obtain the optimum operating conditions.

[Functions Related to the Output Signal]
The output signal can be held at a value immediately before calibration or at a desired value dur-

ing calibration or at the setting/service level. This is to prevent the output signal from adversely

affecting the operation of the equipment configuring the measurement loop (default setting: the

output signal is held at the value immediately before).

[Functions Related to the High/Low Limit Alarm Contact Output]
Select the parameters for which the alarms are raised, and then set their respective high/low limit

setpoints. The analyzer transmits an alarm contact signal if any of these setpoints is exceeded.

(Define the voltage-free contacts and the state of contacts at the service level.)

[Functions Related to the FAIL Contact Signal]
If a FAIL state occurs, an appropriate message appears on the LCD display. For details on these

functions, see Chapter 11.

[Contact output of Hydrogen purity / Replacement meter]
According to switching of Replacement meter, the status of contact and display (LED) is changed

as following. (The status of contact output are to be set in Service level, CODE 05.)

• Hydrogen purity meter

RANGE

FUNC

SEL GAS

LED

Replacement

Meter

H

2

in CO

2

AIR in CO

2

ACTIVE

NORMAL

ON

OFF

Hydrogen Purity/

Replacement Meter

H

2

in AIR

H

2

in CO

2

AIR in CO

2

NORMAL

ACTIVE

ACTIVE

NORMAL

ACTIVE

NORMAL

OFF

ON

OFF

FUNC: Identification signal of Hydrogen purity/Replacement Meter. (contact)

SEL GAS: Range identification signal of Replacement Meter. (contact)

[The status of contact output]

FAIL

SEL

FUNC

MAINT

ALM

Power Supply

OFF

OPEN

CLOSE

CLOSE

CLOSE

CLOSE

Power Supply

ON

NORMAL: CLOSE

FAILURE: OPEN

The status of contact are to be set in Service level, CODE 05.