Advertising
background image

<1. Specifications>

1-6

IM 11T3E1-01E

l

Contact Outputs/Input

Contact output:

Signals for Maintenance, Fail, Hi/Lo alarms

Contact capacity: 250 V AC at 3 A or 30 V DC at 3 A

Contact input:

Signal for switching between the Hydrogen Purity meter and the Replace-

ment meter

l

Calibration

Manual (one touch), Semi automatic, Automatic calibration

l

Communication

Protocol:

BRAIN communication

Data items that can be transmitted by the hand-held terminal are

numerical data, such as concentration, temperature and pressure,

alarm setpoint and self-diagnostic parameters.

l

Ambient Conditions

Temperature:

-10 to 55 °C (14 to 131°F)

Humidity:

5 to 95 %RH

l

Installation

Non-explosion-proof models: Pipe-, panel- or wall-mounted
Explosion-proof models:

Pipe- or panel-mounted

l

Finish

GD402G (general purpose):
Front cover:

Equivalent to Munsell 0.6GY3.1/2.0

Case:

Equivalent to Munsell 2.5Y8.4/1.2

GD402T, V, R (explosion-proof):

Equivalent to Munsell 0.6GY3.1/2.0

l

Weight

GD402G (general purpose): Approx. 3 kg (including mounting hardware)
GD402T, V, R (explosion-proof): Approx. 15 kg (including mounting hardware)

l

Fuse

250 V 1 A quick acting type authorized VDE/SEMKO 100 to 240 V AC Model
250 V 2 A quick acting type authorized VDE/SEMKO 24 V DC Model