<6. CALORIE / PARAMETER SETTING>

6-3

IM 11T3E1-01E

6.1.2

Setting Parameters at Operation Level

Press the [MODE] key and enter the password (XXX) to gain access to this level (see Figure 6.1

for key operation).

Table 6.2 Setting Parameters at Operation Level

Mode/Setting parameter

Display

Data to be Set

(or Conditions)

Remarks

Calibration mode

Semi-automatic calibration

Manual calibration

Setting of zero-point calorific

value

Setting of zero-point density

Setting of span-point calorific

value

Setting of span-point density

SEM.CAL

MAN.CAL

Z_CAL

Z_DNS

S_CAL

S_DNS

One-touch calibration

See Figure 9.3 for key operation.

See Figure 9.4 for key operation.

Zero/span-gas valve operation

Zero-gas valve operation

Span-gas valve operation

Sample-gas valve operation

VALVE

No setting is required for

manual

calibration.

Control of zero-gas valve

Control of span-gas

valve

Switching between sam-

ple and

calibration gases

See Figure 9.1 for key operation.

Density/calorie conversion factor

setting mode

Automatic zero-point adjustment

Factor setting

Zero point

Factor K0

Factor K1

ADJUST

AT_ADJ

SET_K

K0_ADJ

K0

K1

The decimal point can

be repositioned for this

setting only.

Range : -99999 to 99999

Range : -99999 to 99999

Range : 0 to 99999

See Figure 6.2 for key operation.

(Floating-point)