<6. CALORIE / PARAMETER SETTING>

6-29

IM 11T3E1-01E

CODE 40: Indication of Calibration Coefficients (Read-only)

This setting determines the calibration coefficients to be indicated.

(a) Zero [*C_K_Z]:

indicated as SX.XXXX

(b) Span [*C_K_S]:

indicated as X.XXXX

[ENT]

40

*CODE

*SERVC

SX.XXXX

*C_K_Z

X.XXXX

*C_K_S

[ENT]

[ENT]

[YES]

F0624.ai

*

1

*

2

*

3

*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.

*2: Enter the code number.

*3: By pressing the [ENT] key, the next indication is displayed.

This parameter does not accept the [ ] and [ ] key input.

Figure 6.24 Indication of Calibration Coefficients

CODE 41: Indication of Oscillation Frequency (Read-only)

This setting determines the F2 and F4 oscillation frequencies coefficients to be indicated.

(a) Indication of frequency F2 (kHz) [*F2.KHZ]:

indicated as X.XXXXX

(b) Indication of frequency F4 (kHz) [*F4.KHZ]:

indicated as X.XXXXX

(c) Indication of frequency ratio (kHz) [*F2/F4]:

indicated as X.XXXXX

[ENT]

41

*CODE

*SERVC

X.XXXXX

*F2.KHZ

X.XXXXX

*F4.KHZ

[ENT]

[ENT]

[YES]

F0625.ai

*

1

*

2

*

3

*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.

*2: Enter the code number.

*3: By pressing the [ENT] key, the next indication is displayed.

This parameter does not accept the [ ] and [ ] key input.

[ENT]

X.XXXXX

*F2/F4

*

3

*

3

Figure 6.25 Indication of Oscillation Frequency