<2. Installation, Wiring and Piping>

2-3

IM 11T3E1-01E

2.1.3

GD40V (CSA Explosion-proof and Intrinsically Safe

Approval)

Sealing Fitting

Conduit

500 mm (20") Max.

NON-HAZARDOUS

LOCATIONS

HAZARDOUS

LOCATIONS

GD40V

F0203.ai

Figure 2.2

1. GD40V Gas density detector for use in hazardous (classified) locations:

• Explosion-proof for Class I, Division 1, Groups B, C and D;

Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G with Intrinsically

Safe sensor for Class I, II, III, Division 1, Groups B, C, D, E, F and G.

• Enclosure Rating:

Type 4X

• Temperature Code:

T5

• Ambient Temperature: -10 to 60°C

2. Wiring

• All wiring shall comply with Canadian Electrical and Local Electrical codes.

• Note a warning label worded as follows.

WARNING

A seal shall be installed within 500 mm (20 inches) of the enclosure. Refer to Figure 2.2.

3. Operation

• Note a warning label worded as follows.

WARNING

• Open circuit before removing cover.
• Substitution of components may impair instrinsic safety.

• Take care not to generate mechanical spark when access to the instrument and

peripheral devices in hazardous locations.

4. Maintenance and Repair

• The instrument modification or parts replacement by other than authorized representative

of Yokogawa Electric Corporation is prohibited and will void the certification of CSA

International.