Yokogawa GD40 Gas Density Detector User Manual

Page 11

Advertising
background image

<1. Specifications>

1-2

IM 11T3E1-01E

(2) GD402G, T, V, R Converter : Rainproof for outdoor use (equivalent to IP65 / NEMA 4X)

Ambient Temperature :

-10 to 55°C

Ambient Humidity :

5 to 95%RH

GD402G : General purpose converter. (Non Explosion-proof).

Electrical connection : 21mm (0.9inch) in diameter. Pg 13.5 cable glands includ-

ed

GD402T : FM Explosion-proof Approval.

Explosion-proof for Class I, Division 1, Groups B, C and D;

Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.

Enclosure :

NEMA Type 4X

Temperature Code : T6

Electrical connection : 1/2 NPT female

GD402V : CSA Explosion-proof Approval.

Explosion-proof for Class I, Division 1, Groups B, C and D;

Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.

Enclosure :

Type 4X

Temperature Code : T6

Electrical connection : 1/2 NPT female

GD402R : TIIS Explosion-proof Approval.

Explosion-proof code : Exd IIB+H

2

T6

Temperature Code : T6

Electrical connection : G3/4 female

(3) EJA310 Absolute pressure transmitter (optional)

FM Explosion-proof Approval:

Explosion-proof for Class I, Division 1, Groups B, C and D;

Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.

Hazardous(classified locations, indoors and outdoors (NEMA 4X)

Temperature Code : T6

Ambient Temperature : -40 to 60°C

Ambient Humidity :

5 to 100%RH (at 40°C)

Electrical connection : 1/2 NPT female

Process connection : 1/4 NPT female

CSA Explosion-proof Approval:

Explosion-proof for Class I, Division 1, Groups B, C and D;

Dust Ignition-proof for Class II, III, Division 1, Groups E, F and G.

Division2 ‘SEALS NOT REQUIRED’

Enclosure :

Type 4X

Temperature Code : T4, T5, T6

Max. Process Temp.: T4 120°C, T5 100°C, T6 85°C

Ambient Temperature : -40 to 80°C