<5. DENSITY / PARAMETER SETTING>

5-8

IM 11T3E1-01E

(2) Display Mode
Select one of the following ten parameters to be shown in the message field. Each press of the

[NO] key cycles through the choices in the order shown below:

1. Physical density [XX.XXXX KG/M3] or [X.XXXXX LB/FT3]

(depends on the setting of CODE 21)

2. Compensated density [X.XXXX KG/M3 ]or [X.XXXXX LB/FT3]

(depends on the setting of CODE 21)

3. Specific gravity [X.XXXX SP GR]

4. Calorific value [XXX.XXX MJ/M3] or [X.XXXX KBTU]

(depends on the setting of CODE 22)

5. Molecular weight [XXX.XX MOL]

6. Concentration[XXX.X VOL% ]

7. Temperature [XXX.X °C] or [XXX.X °F]

(depends on the setting of CODE 23)

8. Pressure [XXX.XX KPA] or [X.XXXX MPA] or [XX.XXX PSI]

(depends on the setting of CODE 20)

9. Current for analog output 1 (%) [XXX.X MA1%]

10. Current for analog output 2(%) [XXX.X MA 2%]

Show the desired choice and press the [YES] key. That choice is set as a parameter to be

measured in the measurement mode (<MEASURE>). Parameters 1 to 6 are configurable, while

parameters 7 to 10 are read-only.