Yokogawa GD40 Gas Density Detector User Manual

Page 155

Advertising
background image

<8. DENSITY / CALIBRATION PROCEDURE>

8-2

IM 11T3E1-01E

(2) Operations for Setting the Automatic/Remote Semi-automatic Calibration
• Selection for automatic calibration [*AUTO.C]

At the service level, select CODE 13.
• Disable: 0
• Enable: 1

• Selection for remote semi-automatic calibration [*REMOT]

At the service level, select CODE 14.
• Disable: 0
• Enable: 1

(3) Valve Check
Check the performance of valves according to the “valve operation flowchart” at the operation

level.

[NO]

VALVE

[NO]

XXX.XXX

V_ZERO

MAN.CAL

END

XXX.XXX

Z_OPEN

XXX.XXX

V_SPAN

XXX.XXX

S_OPEN

(Zero-gas valve operation)

[YES]

[YES]

[YES]

[YES]

or

[NO]

[NO]

[NO]

[NO]

WAIT

[YES]

*2

Measurement mode

*3

[NO]

[NO]

*7

*7

*7

*7

*2

*4

*4

*6

*5

*1

(Span-gas valve operation)

(Stabilization time)

*1: See "Figure 5.1 Modes of Operation Level" on page 5-8.

*2: Press the [NO] key to return to the parameter for sample gas.

*3: Using the [NO] key, show [V_ZERO], [V_SPAN] and [END] in succession.

*4: Using the [YES] key, open the zero-gas or span-gas value to read the measured value

at the point.

*5: Press the [YES] key to go into the measurement mode state after a given stabilization

time elapses; press the [NO] key to go into that state immediately.

*6: This step is skipped if no valve was opened.

*7: The display shows the same data item to be measured as the one in the measurement mode.

F0802.ai

DISP

Figure 8.1 Valve Operation Mode

Advertising