Code 10: setting of pressure compensation – Yokogawa GD40 Gas Density Detector User Manual

Page 143

Advertising
background image

<7. HYDROGEN PURITY / PARAMETER SETTING>

7-20

IM 11T3E1-01E

CODE 10: Setting of Pressure Compensation

This setting determines whether or not the measured density should be compensated by the gas

pressure. The density can be compensated either by receiving a signal from a pressure transmit-

ter or by using a fixed pressure setpoint.
If you want the analyzer to compensate the density using a pressure transmitter, you MUST

connect a pressure transmitter to the converter and enter the pressure input and zero and span

points with CODE 12 at the service level.
If you do not want the analyzer to compensate the density, the analyzer shows and outputs the

measured density equivalent to a value at an atmospheric pressure of unity (101.33 kPa ABS)..

(a) Enable/disable/select preset value [*P.COMP]
・Disable:

0

・Enable (value immediately before): 1
・Enable (fixed value):

2

(b) Setting of fixed value for pressure compensation [*P.FIX]
This item applies if item (a) above is set to "2".
Configurable range: 0.10 to 999.99 (for kPa),

0.0001 to 9.9999 (for MPa),

0.015 to 99.999 (for psi )

[ENT]

10

*CODE

*SERVC

X

*P.COMP

XXX.XX

*P.FIX

[ENT]

[ENT]

[YES]

F0713.ai

*

1

*

2

*

3

*

4

*

5

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: Set a value using the [ ] and [ENT] keys.

*4: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].

*6: The position of the decimal point changes depending on the setting of CODE 20.

If the code is set to "0" (kPa), the reading is XXX.XX;

if the code is set to "1" (MPa), the reading is X.XXXX;

if the code is set to "2" (psi), the reading is XX.XXX.

*

6

Figure 7.13 Setting of Pressure Compensation

Advertising
This manual is related to the following products: