<5. DENSITY / PARAMETER SETTING>

5-11

IM 11T3E1-01E

1. Analog Output Setting Mode [ *RANGE ]
The analog output has output 1 and output 2. Only output 1 can be used for communication

purposes.
Set a range appropriate for 4-20 mA DC output signals. The range has two setpoints: the lower

limit (zero point) of the range corresponding to the minimum (0%) of a given output signal and the

upper limit (span point) of the range corresponding to the maximum (100%) of the given output

signal.
(1) Output1

[*OUT1]

(a) Select a parameter to be output.
・ Physical density [*DENS]
・ Compensated density [*C_DNS]
・ Specific gravity [*SP_GR]
・ Calorific value [*CALRY]
・ Molecular weight [*MOL]
・ Concentration [*CONCT]
・ Gas temperature [*TEMP]
・ Gas pressure [*PRESS]
(b) Set the zero and span points. (This item is not applied if temperature or pressure are cho-

sen as the parameter in item (a) above.)

・ Physical density [*Z_DNS], [*S_DNS]

Configure range:

00.0000 to 60.0000 (kg/m

3

),

0.00000 to 4.00000 (lb/ft

3

)

・ Compensated density [*Z_CP.D], [*S_CP.D]

Configure range:

0.0000 to 6.0000 (kg/m

3

),

0.00000 to 0.40000 (lb/ft

3

)

・ Specific gravity [*Z_SPC], [*S_SPC]

Configure range:

0.0000 to 5.0000

・ Calorific value [*Z_CAL], [*S_CAL]

Configure range:

000.000 to 133.000(MJ/m

3

),

0.0000 to 3.5000 (kBTU/ft

3

)

・ Molecular weight [*Z_MOL], [*S_MOL]

Configure range:

000.00 to 140.00

・ Concentration [*Z_CON], [*S_CON]

Configure range:

000.0 to 100.0

(2) Output2

[*OUT2]

The same parameters as those of item (1) apply.