<5. DENSITY / PARAMETER SETTING>

5-28

IM 11T3E1-01E

CODE 30: Setting of Calendar

This setting defines the date and time.

(a) Setting of year/month/day [*Y_M_D]
Configurable range : 00.01.01 to 99.12.31
(b) Setting of hour/minute [*H_M]
Configurable range : 00.00 to 23.59

30

*CODE

*SERVC

XX.XX.XX

*Y_M_D

XX.XX

*H_M

F0522.ai

*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

[ENT]

[ENT]

[YES]

[ENT]

*

3

*

3

*

2

*

1

Figure 5.21 Setting of Calendar

CODE 31: Showing/Hiding of Negative Measured Values [*MINUS]

This setting determines whether negative measured values (-) should be shown or hidden.

・ Show: 0
・ Hide: 1

[ENT]

*CODE

31

*SERVC

X

*MINUS

[ENT]

[YES]

F0523.ai

*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

*

3

*

2

*

1

Figure 5.22 Showing/Hiding of Negative Measured Values