Yokogawa GD40 Gas Density Detector User Manual

Page 48

Advertising
background image

<3. Operation>

3-5

IM 11T3E1-01E

3. Entering Data Values

NOTE

The data you have entered keep in memory even if you turn off the power. If the data you have

entered are provisional, reenter the normal data.

The following explains how to make changes to existing data entries.
In the display shown in Figure 3.4-(2), pressing the [YES] key causes the message field to default

to [*DENS]. Pressing the [YES] key once again changes the message to [*Z_DNS], the range of

lower limit setting, and makes the indication of the lower limit blink.
Assume the existing data entry of the range's lower limit is 1.2927. To change the value to

2.0535, follow the instructions given below:
(1) Press the [>] key so “1“ starts blinking.
(2) Press the

[ ]

>

key until the blinking "1" changes to ''2".

(3) Press the [>] key until the next digit begins blinking. Then, press the

[ ]

>

key until the blink-

ing ”2” digit changes to “0".

(4) Press the [>] key until the next digit begins blinking. Then, press the

[ ]

>

key until the blinking

digit changes to ''5”.

(5) Press the [>] key until the next digit begins blinking. Then, press the

[ ]

>

key until the blinking

digit changes to ''3”.

(6) Press the [>] key until the next digit begins blinking. Then, press the

[ ]

>

key until the blink-

ing digit changes to ''5”.

(7) Press the [ENT] key. This confirms the value “2.0535".

Pressing the

[ENT] key enters the value shown

in the data field.

(1)

(2)

F0305.ai

In the display shown in Figure 3.4 (2), pressing the

[YES] key will change the display to look like this.

Figure 3.5 Display During Data Setting