<7. HYDROGEN PURITY / PARAMETER SETTING>

7-25

IM 11T3E1-01E

CODE 31: Showing/Hiding of Negative Measured Values [*MINUS]

This setting determines whether negative measured values (-) should be shown or hidden.

・ Show: 0
・ Hide: 1

[ENT]

*CODE

31

*SERVC

X

*MINUS

[ENT]

[YES]

F0720.ai

*

1

*

2

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

*

3

Figure 7.20 Showing/Hiding of Negative Measured Values

CODE 40: Indication of Calibration Coefficients (Read-only)

This setting determines the calibration coefficients to be indicated.

(a) Zero [*C_K_Z]:

indicated as SX.XXXX

(b) Span [*C_K_S]:

indicated as X.XXXX

[ENT]

40

*CODE

*SERVC

SX.XXXX

*C_K_Z

X.XXXX

*C_K_S

[ENT]

[ENT]

[YES]

F0721.ai

*

1

*

2

*

3

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: By pressing the [ENT] key, the next indication is displayed.

This parameter does not accept the [ ] and [ ] key input.

Figure 7.21 Indication of Calibration Coefficients