<2. Installation, Wiring and Piping>

2-5

IM 11T3E1-01E

2.2

Installing the Converter

The GD402 converter comes in either the non-explosion-protected Model GD402G or explosion-

protected Model GD402T, V, R (explosion-protection code: Refer to 1.1.1).

Choose the Model GD402T or V or R if the converter is to be used in a hazardous area.

2.2.1

Selecting the Location

l

Ease of operation

Select a location where you can easily view the readings on the display and work with the keys.

Installing the converter closer to the detector will ease your maintenance work, including calibra-

tion.

l

Explosion-proof construction

When installing the converter in an explosion-hazardous area, ensure that the area conforms to

the explosion-proof code noted above.

l

No corrosive gases

Corrosive gases are not desirable because they may damage the electrical components in the

converter.

l

Slight mechanical vibration

Although the converter is vibration-resistant, vibration may loosen the connections of the external

wiring.

l

No exposure to direct sunlight

Exposure to direct sunlight may raise the temperature in the converter to abnormal levels, and

should therefore be avoided. Note that such abnormal temperature levels can also result from

heat irradiated from high-temperature equipment around the converter. If the converter is likely to

receive direct sunlight and the temperature in the converter will likely exceed the operating limits,

attach a hood (optional).

l

Humidity maintained between 5% and 95% RH.

Avoid choosing a location that is likely to be exposed to abnormally high or low humidity over a

prolonged period. It is recommended that the converter be used at a humidity between 25% and

85 % RH.

l

No exposure to rainwater

Even though the converter is rainproof, whenever possible install it where it is protected from

water splashes. The reasoning for this is that the converter cover may need to be removed for

maintenance or for other reasons

l

Altitude of installation site is lower than 2,000 m.