Code 30: setting of calendar – Yokogawa GD40 Gas Density Detector User Manual

Page 117

Advertising
background image

<6. CALORIE / PARAMETER SETTING>

6-28

IM 11T3E1-01E

CODE 30: Setting of Calendar

This setting defines the date and time.

(a) Setting of year/month/day [*Y_M_D]
Configurable range : 00.01.01 to 99.12.31
(b) Setting of hour/minute [*H_M]
Configurable range : 00.00 to 23.59

[ENT]

30

*CODE

*SERVC

XX.XX.XX

*Y_M_D

XX.XX

*H_M

[ENT]

[ENT]

[YES]

F0622.ai

*

1

*

2

*

3

*

3

*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

Figure 6.22 Setting of Calendar

CODE 31: Showing/Hiding of Negative Measured Values [*MINUS]

This setting determines whether negative measured values (-) should be shown or hidden.

・ Show: 0
・ Hide: 1

[ENT]

*CODE

31

*SERVC

X

*MINUS

[ENT]

[YES]

F0623.ai

*

1

*

2

*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

*

3

Figure 6.23 Showing/Hiding of Negative Measured Values