<5. DENSITY / PARAMETER SETTING>

5-19

IM 11T3E1-01E

CODE 05: Setting of Contact Outputs Status [*CNTCT]

This setting defines the status of the contact outputs.

Configurable range: 00 to 15

05

*CODE

*SERVC

XX

*CNTCT

[YES]

F0511.ai

*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

[ENT]

[ENT]

*

3

*

2

*

1

Figure 5.10 Setting of Contact Outputs Status

Each contact output takes either of the following two status (in table), depending on the value you

set.

Value

ZERO contact

SPAN contact MAINT contact

ALM contact

0

NO

NO

NO

NO

1

NO

NO

NO

NC

2

NO

NO

NC

NO

3

NO

NO

NC

NC

4

NO

NC

NO

NO

5

NO

NC

NO

NC

6

NO

NC

NC

NO

7

NO

NC

NC

NC

8

NC

NO

NO

NO

9

NC

NO

NO

NC

10

NC

NO

NC

NO

11

NC

NO

NC

NC

12

NC

NC

NO

NO

13

NC

NC

NO

NC

14

NC

NC

NC

NO

15

NC

NC

NC

NC

NO: means “normally open”

NC: means “normally closed”