<5. DENSITY / PARAMETER SETTING>

5-24

IM 11T3E1-01E

CODE 15: Setting of Data for Automatic/Semi-automatic Calibration

This setting defines the data needed to carry out automatic/semi-automatic calibration.

(a) Selection of calibration item [*CAL.P]
・ Zero and span:

0

・ Zero:

1

・ Span:

2

(b) Setting of calibration time [*CAL.T]

Configurable range: 00 to 60 (unit : minute)

(c) Setting of stabilization time [*STAB.T]

Configurable range: 00 to 60 (unit : minute)

(d) Setting of calibration starting time
(1) Setting of year/month/day [*Y_M_D]

Configurable range: 00.01.01 to 99.12.31

(2) Setting of hour/minute [*H_M]

Configurable range: 00.00 to 23.59

(e) Selection of unit of calibration interval [*CYCL.U]
・ Hour: 0
・ Day: 1
(f) Selection of calibration interval [*CYCL.T]
Item (e) above is set to "0".

Configurable range: 00 to 23 (for hour)

Item (e) above is set to "1".

Configurable range: 000 to 255 (for day)

For more information, see Chapter 8, “Calibration Procedure”.