2 setting calibration data, Setting calibration data -3 – Yokogawa GD40 Gas Density Detector User Manual

Page 156

Advertising
background image

<8. DENSITY / CALIBRATION PROCEDURE>

8-3

IM 11T3E1-01E

8.2.2

Setting Calibration Data

To set calibration data, press the [*] key in the measurement mode, (when password is selected,

enter the password (XXX), ) and select the parameter [*CAL.DT]. See Figure 5.4 for the process

at the setting level.

(a) Zero-point density [*Z_DNS]

Configurable range : 0.0000 to 6.0000 (kg/m

3

)

0.00000 to 0.40000 (lb/ft

3

)

(b) Span-point density [*S_DNS]

Configurable range : 0.0000 to 6.0000 (kg/m

3

)

0.00000 to 0.40000 (lb/ft

3

)

(c) Output hold value during calibration [*C_HLD]
• Disable:

0

• Enable:

1

• Enable (preset value):

2

(d) Setting of preset value [*PR.SET]

Configurable range:

-10.0 to 110.0

(Applies only if item (c) above is set to “2”.)

 

[NO]

[ENT]

*RANGE

[NO]

X.XXXX

*Z_DNS

*ALARM

*CAL.DT

X.XXXX

*S_DNS

X

*C_HLD

XXX.X

*PR.SET

[ENT]

[ENT]

[ENT]

F0803.ai

[YES]

*

2

*

2

*

2

*

2

*

3

*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.

*2: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

*3: This step is skipped if the parameter [*C_HLD] is set to a value other than "2".

*

1

(Setting of Zero-point density)

(Setting of preset value)

(Setting of Output hold value during calibration)

(Setting of Span-point density)

Figure 8.2 Calibration Parameter Setting Mode