<10. HYDROGEN PURITY / CALIBRATION PROCEDURE>

10-2

IM 11T3E1-01E

10.3 Calibration

In the manual calibration mode, visually check that the reading has settled, and then confirm the

reading manually (by pressing the [ENT] key).

[NO]

[NO]

0.0899

H2

MAN.CAL

[YES]

*

1

ALM.10

ZERO

[YES]

Measurement mode

*

2

*

5

*

3

[NO]

1.9971

CO2

[YES]

*

5

SPAN

*

4

X.XXXX

CAL.SET

X.XXXX

CAL.SET

[ENT]

[ENT]

[ENT]

[ENT]

*

4

*

5

*

3

[NO]

*1: See "Figure 7.2 Modes of Operational Level" on page 7-7.

*2: If a calibration data error occurs, the display shows ALM.10 .

Pressing the [YES] or [NO] key returns this status to the [MAN.CAL].

*3: The displayed value is fixed. It cannot be set.

*4: When the reading settles, press the [ENT] key to confirm your calibration.

This parameter does not accept input with the [ ] and [ ] keys.

The display always shows the corrected density.

*5: Pressing the [NO] key for both the [ZERO] and [SPAN] indications returns

this status to the [MAN.CAL].

Zero-point density (H

2

gas) is displayed.

Close the zero gas valve.

Close the sample gas valve.

Span-point density (CO

2

gas) is displayed.

Close the span gas valve.

Open the sample gas valve.

Open the zero gas valve.

Press [ENT] key after the value is stabilized.

Open the span gas valve.

Press [ENT] key after the value is stabilized.

S_GAS

F1003.ai

DISP

Figure 10.2 Manual Calibration Mode