<8. DENSITY / CALIBRATION PROCEDURE>

8-1

IM 11T3E1-01E

8.

DENSITY /CALIBRATION

PROCEDURE

This chapter describes how to calibrate of Density meter.

If select Calorie meter, see Chapter 9.

If select Hydrogen purity meter, see Chapter 10.
Zero and span calibration can be carried out using standard gases (zero and span gases). Three

modes are available for calibration: automatic, semi-automatic (remote semi-automatic) and

manual. Even if automatic calibration is enabled in the parameter setting, you can interrupt the

system with either the semi-automatic or manual mode of calibration. The following explains the

basic procedure and operation used in calibration.

8.1

Basic Calibration Procedure

Semi-automatic calibration mode

Calibration parameter setting

See the process in Figure 8.2

See the process in Figure 5.1

Selection of semi-automatic/manual calibration

See the process in Figure 8.4

See the process in Figure 8.3.

Manual calibration mode

F0801.ai

8.2

Preparation for Calibration

Prepare for calibration during the initial operation of the analyzer, by carrying out the following.

• At the operation level, set the parameters for semi-automatic and manual calibration. Also

set the valve-operation parameter for switching between gases to check the valve perform-

operation parameter for switching between gases to check the valve perform-

peration parameter for switching between gases to check the valve perform-

ance.

• At the setting level, set the calibration data (zero- and span-point values, status of output

during calibration etc.)

• At the service level, determine whether automatic and remote semi-automatic calibrations

are enabled or disabled in CODE 13 and 14. Set the calibration data (selection of calibration

item, calibration time, stabilization time, etc.) in CODE 15.

8.2.1

Setting Type of Calibration and Checking Valve

Operation

(1) Operations to Set the Semi-automatic and Manual Calibration
In the measurement mode, press the [MODE] key, (if password is selected, enter the password

(XXX) (See section “4.2 Setting Lists”), ) and select the desired type of calibration.

• To select semi-automatic calibration, show [SEM.CAL] and press the [YES] key.
• To select manual calibration, show [MAN.CAL] and press the [YES] key.
For details on this process, see Figure 5.1.