2 parameter setting at setting level, Parameter setting at setting level -10 – Yokogawa GD40 Gas Density Detector User Manual

Page 65

Advertising
background image

<5. DENSITY / PARAMETER SETTING>

5-10

IM 11T3E1-01E

5.2.2

Parameter Setting at Setting Level

To move to the setting level, press the [

*

] key and then enter the password (XXX). The main task

at the setting level is to set data values such as the measuring range values.
Note that the function for which you have set data values at the setting level, will not work if that

function is turned off at another level. For this reason, care must be taken when dealing with

modes that relate to each other. The setting level has four levels/modes, as shown below, al-

though this subsection explains only the analog output setting and alarm-point setting modes.

The calibration parameter setting mode is discussed in Chapter 8.

1. Analog output setting mode [*RANGE]
2. Calibration parameter setting mode [*CAL.DT]
3. Alarm-point setting mode [*ALARM]
4. Service level [*SERVC]

Select a mode from the above four choices. Each press of the [NO] key cycles through the

choices in sequence. Press the [YES] key to enter the desired mode when it is indicated.

You can abort the setting operation in any of the modes by pressing the [MODE] key.

Doing this will return the analyzer to the measurement mode.

Password is not selected in first condition.

(See "Figure 5.27 Process of Setting Password" on page 5-31.)

[ENT]

[NO]

[NO]

[NO]

[YES]

[YES]

[ENT]

[ENT]

[ENT]

[YES]

[

*

]

Analog output setting mode

Calibration parameter setting mode

Alarm-point setting mode

Measurement Mode

*RANGE

*CAL.DT

*ALARM

See the process in Figure 5.4.

See the process in Figure 8.2.

See the process in Figure 5.5.

F0503.ai

[YES]

[ENT]

Service level

See the process in Figure 5.7 to 5.30.

*SERVC

[NO]

XXX

*PASSW

Figure 5.3 Process at the Setting Level