Yokogawa GD40 Gas Density Detector User Manual

Page 80

Advertising
background image

<5. DENSITY / PARAMETER SETTING>

5-25

IM 11T3E1-01E

X

*CAL.P

(Selection of calibration item)

XX

*CAL.T

(Setting of calibration time)

XX

*STAB.T

(Setting of stabilization time)

XX.XX.XX

*Y_M_D

(Setting of calibration starting time [year/ month/ day])

XX.XX

*H_M

(Setting of calibration starting time [hour/ minute])

X

*CYCL.U

(Selection of unit of calibration interval)

XXX

*CYCL.T

(Selection of calibration interval)

[ENT]

*SERVC

[ENT]

[YES]

F0517.ai

*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.

*2: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

*3: The number of digits on LCD varies depending on the unit of the calibration interval (parameter [*CYCL.U]).

If the parameter is set to "0" (hour), the indication is two-digit number.

If the parameter is set to "1" (day), the indication is three-digit number.

15

*CODE

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

[ENT]

*

1

*

2

*

2

*

2

*

2

*

2

*

2

*

2

*

2

*

3

Figure 5.16 Setting of Data for Automatic/Semi-automatic Calibration