Yokogawa GD40 Gas Density Detector User Manual

Page 141

Advertising
background image

<7. HYDROGEN PURITY / PARAMETER SETTING>

7-18

IM 11T3E1-01E

CODE 04: Setting of Output-smoothing Constants [*SMOTH]

This setting defines the constants for output smoothing.

This smoothing constants is output-smoothing constants of converter (electrical circuit boards).

Configurable range: 00 to 60. (unit : second)

[ENT]

04

*CODE

*SERVC

XX

*SMOTH

[ENT]

[YES]

F0711.ai

*

1

*

2

*

3

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

Figure 7.11 Setting of Output-smoothing Constants

CODE 05: Setting of Contact Outputs Status [*CNTCT]

This setting defines the status of the contact outputs.

Configurable range: 00 to 15

[ENT]

05

*CODE

*SERVC

XX

*CNTCT

[ENT]

[YES]

*

1

*

2

*

3

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

F0712.ai

Figure 7.12 Setting of Contact Outputs Status

Advertising