Yokogawa GD40 Gas Density Detector User Manual

Page 12

Advertising
background image

<1. Specifications>

1-3

IM 11T3E1-01E

Ambient Humidity :

5 to 100%RH (at 40°C)

Electrical connection : 1/2 NPT female

Process connection : 1/4 NPT female

TIIS Explosion-proof Approval:

Explosion-proof code : Ex do IIC T4X

Temperature Code : T4

Ambient Temperature : -20 to 60 °C

Ambient Humidity :

5 to 100%RH (at 40°C)

Electrical connection : G1/2 female

Process connection : Rc1/4

l

Sample Gas conditions

Sample gas:

All gases except for corrosive gas and acetylene gas.

Temperature:

-10 to 50 °C (non-condensing)

Pressure:

Max. 588.5 kPa (abs)

Gas flow:

0.1 to 1 l/min

l

Output Signals

Output 1: 4-20 mA DC

Isolated from inputs; load resistance: 600 W maximum (Load resistance of 250-

550 W required when in the BRAIN communication mode.)

Output 2: 4-20 mA DC

Isolated from inputs; load resistance: 600 W maximum

l

Power Supply

Rated voltage range:

100 to 240 V AC, 24 V DC

Allowable voltage range:

85 to 264 V AC, 21.6 to 26.4 V DC

Rated frequency:

50 or 60 Hz

Allowable frequency range: 47 to 63 Hz

l

Power Consumption

Approximately 12 W.

l

Safety and EMC Compliance

Safety Standards: EN61010-1
EMC Standards:
(Applied only when GD402G converter is used with GD40G detector.)
Emission:

EN61326 Class A

EN61000-3-2

EN61000-3-3

AS/NZS CISPR11

Immunity:

EN61326 Annex A and F

KC Marking:

Korea Electromagnetic Conformity Standard