<7. HYDROGEN PURITY / PARAMETER SETTING>

7-26

IM 11T3E1-01E

CODE 41: Indication of Oscillation Frequency (Read-only)

This setting determines the F2 and F4 oscillation frequencies coefficients to be indicated.

(a) Indication of frequency F2 (kHz) [*F2.KHZ]:

indicated as X.XXXXX

(b) Indication of frequency F4 (kHz) [*F4.KHZ]:

indicated as X.XXXXX

(c) Indication of frequency ratio (kHz) [*F2/F4]:

indicated as X.XXXXX

[ENT]

41

*CODE

*SERVC

X.XXXXX

*F2.KHZ

X.XXXXX

*F4.KHZ

[ENT]

[ENT]

[YES]

F0722.ai

*

1

*

2

*

3

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: By pressing the [ENT] key, the next indication is displayed.

This parameter does not accept the [ ] and [ ] key input.

[ENT]

X.XXXXX

*F2/F4

*

3

*

3

Figure 7.22 Indication of Oscillation Frequency

CODE 42: Indication of Software Version [*REV] (Read-only)

This setting allows the software version to be confirmed (indicated).
Indicated as X.XX

42

*CODE

*SERVC

X.XX

*REV

[ENT]

[ENT]

[YES]

F0723.ai

*

1

*

2

*

3

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: Pressing the [ENT] key returns to the service level with the indication [*SERVC].

This parameter does not accept the [ ] and [ ] key input.

Figure 7.23 Indication of Software Version