<7. HYDROGEN PURITY / PARAMETER SETTING>

7-13

IM 11T3E1-01E

2. Alarm-point Setting Mode [*ALARM]
Set the upper /lower limits for the measured values of the parameters noted below, for the pur-

pose of raising an alarm.

(a) Select a parameter for which alarms are raised.
・ Physical density [*DENS]
・ Compensated density [*C_DNS]
・ Hydrogen purity [*H_A]
Press the [YES] key to enter the desired mode when it is indicated.

(b) Set the high/low limits.
・ Physical density [*L_DNS], [*H_DNS]

Configure range:

00.0000 to 60.0000 (kg/m

3

),

0.00000 to 4.00000 (lb/ft

3

)

・ Compensated density [*L_CP.D], [*H_CP.D]

Configure range:

0.0000 to 6.0000 (kg/m

3

),

0.00000 to 0.40000 (lb/ft

3

)

・ Hydrogen purity [*L_H_A]

Configure range:

0.0 to 100.0

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: The display shows the previous parameter first.

*3: This step is skipped when the replacement meter is selected.

[NO]

[ENT]

[ENT]

*ALARM

[NO]

XX.XXXX

*L_DNS

*CAL.DT

*SERVC

*DENS

XX.XXXX

*H_DNS

X.XXXX

*L_CP.D

X.XXXX

*H_CP.D

XXX.X

*L_H_A

(Physical density)

(Compensated density)

(Hydrogen purity)

*H_A

[ENT]

[ENT]

[ENT]

[YES]

[YES]

[YES]

[NO]

[NO]

[NO]

*C_DNS

F0706.ai

[YES]

*

1

*

2

*

3

*RANGE

[NO]

Figure 7.6 Process of the Alarm-point Setting Mode