11.2.2 Inspecting the Analyzer in a FAIL Status

IN GD402, the state of terminals 18 and 19 is changed from close to open if a FAIL status is

detected by the self-diagnostic function during measurement. In addition, the <FAIL> lamp on

the operation panel comes on and the message field shows an error number. If a FAIL status is

detected, take the corrective action given in Table 11.2.

Table 11.2 Corrective Actions when a FAIL Status Is Detected

Message

Shown

Failure Type

Diagnostic Conditions

Corrective Action

Message via

Communica-

tion

Err.01

Sensor oscillation

shutdown

Failure to detect

oscillation frequency from the

detector

Reset the power

supply.

Contact Yokogawa

service personnel.

FAULT SEN-

SOR

Err.02

Error in sensor’s

oscillation fre-

quency

Frequency F2 : out of

range from 1000 to 10000Hz

Frequency F4 : out of

range from 4000 to 10000Hz

Reset the power

supply.

Contact Yokogawa

service personnel.

OVER FRE-

QUENCY

Err.03

Error in sensor’s

temperature

Sensor’s failure to detect

temperature

Contact Yokogawa

service personnel.

FAULT TEMP

Err.04

A/D converter

failure

Failure in A/D converter

Contact Yokogawa

service personnel.

FAULT A/D

Err.05

Memory failure

Erroneous data in

EEPROM/EPROM/RAM

Contact Yokogawa

service personnel.

FAULT MEMO-

RY