<7. HYDROGEN PURITY / PARAMETER SETTING>

7-5

IM 11T3E1-01E

Table 7.4 Setting Parameters at Service Level

Code

No.

Item

Display

Data to be Set

(or Conditions)

Remarks

40 Indication of calibration coefficients

(read-only)

Zero

Span

*C_K_Z

*C_K_S

SX.XXXX

X.XXXX

See Figure 7.21 for key operation.

41 Indication of oscillation frequency

(read-only)

Frequency F2

Frequency F4

Frequency ratio

*F2.KHZ

*F4.KHZ

*F2/F4

X.XXXXX

X.XXXXX

X.XXXXX

See Figure 7.22 for key operation.

42 Indication of software version

(read-only)

*REV

X.XX

See Figure 7.23 for key operation.

43 Setting of high-resolution mode

*S_CYC

· Normal resolution : 0

· High-resolution : 1

See Figure 7.24 for key operation.

44 Setting of password

*PASS

0.0.0 to 9.9.9

See Figure 7.25 for key operation.

45 Selecting Battery alarm detection

*BAT

· Non detection : 0

· Detection : 1

See Figure 7.26 for key operation.

50 Selecting the instruments

*MODEL

· Density meter : 0

· Calorie meter : 1

· Hydrogen purity, replacement

meter : 2

See Figure 7.27 for key operation.

82 Setting Detector Constant

See Figure 7.28 for key operation.