<5. DENSITY / PARAMETER SETTING>

5-32

IM 11T3E1-01E

CODE 50: Selecting the instruments [ *MODEL]

This setting defines the model.

・ Density meter:

0

・ Calorie meter:

1

・ Hydrogen purity, replacement meter : 2

50

*CODE

*SERVC

X

*MODEL

F0530.ai

*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.

*2: Enter the code number.

*3: Set a value by using the [ ], [ ] and [ENT] keys.

[ENT]

[ENT]

[YES]

*

3

*

2

*

1

Figure 5.29 Selecting the instruments

CODE 82: Setting Detector Constant

This setting defines the detector constant.

CAUTION

The instruments may not only work improperly but also big problems occur unless the detector

constants entered incorrectly. Do carefully and precisely when enter the detector constants into

converter.

1. The GD402 gas density meter is shipped after adjusting the detector and converter in pairs.

When installation, confirm the converter serial number described on the label of detector so

that combines converter and detector correctly. If mismatched in pairs, converter may be out

of order. If combined correctly, no need to enter the detector constants again, as converter

has been adjusted with the constants in factory.

2. When supply converter or detector individually, enter the detector constants, described on

inside the lid of the GD40, into the converter so that the GD402 is going to be well.